18 years of successful growth

Human 25-hydroxyvitamin D-1 alpha hydroxylase, mitochondrial(CYP27B1) ELISA kit

Product Details

Target Name cytochrome P450, family 27, subfamily B, polypeptide 1
Alternative Names 1alpha(OH)ase ELISA Kit; 25-hydroxyvitamin D(3) 1-alpha-hydroxylase ELISA Kit; 25-hydroxyvitamin D-1 alpha hydroxylase ELISA Kit; 25-OHD-1 alpha-hydroxylase ELISA Kit; Calcidiol 1-monooxygenase ELISA Kit; CP27B_HUMAN ELISA Kit; CP2B ELISA Kit; CYP1 ELISA Kit; CYP1ALPHA ELISA Kit; CYP27B ELISA Kit; Cyp27b1 ELISA Kit; Cytochrome p450 27B1 ELISA Kit; Cytochrome p450 27B13 ELISA Kit; Cytochrome P450 family 27 subfamily B polypeptide 1 ELISA Kit; Cytochrome P450 subfamily XXVIIB (25-hydroxyvitamin D-1-alpha-hydroxylase) polypeptide 1 ELISA Kit; Cytochrome P450 subfamily XXVIIB polypeptide 1 ELISA Kit; Cytochrome P450C1 alpha ELISA Kit; Cytochrome P450VD1-alpha ELISA Kit; mitochondrial ELISA Kit; P450C1 alpha ELISA Kit; P450c1 ELISA Kit; P450C1-alpha ELISA Kit; P450VD1-alpha ELISA Kit; PDDR ELISA Kit; VD3 1A hydroxylase ELISA Kit; VDD1 ELISA Kit; VDDR ELISA Kit; VDDR I ELISA Kit; VDDRI ELISA Kit; VDR ELISA Kit
Abbreviation CYP27B1
Species Homo sapiens (Human)
Sample Types serum, plasma, tissue homogenates, cell lysates
Detection Range 31.25 pg/mL-2000 pg/mL
Sensitivity 7.81 pg/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Others
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%   
Three samples of known concentration were tested twenty times on one plate to assess.  
Inter-assay Precision (Precision between assays): CV%   
Three samples of known concentration were tested in twenty assays to assess.    
             
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human CYP27B1 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
  Sample Serum(n=4)  
1:1 Average % 93  
Range % 90-97  
1:2 Average % 89  
Range % 85-96  
1:4 Average % 96  
Range % 90-100  
1:8 Average % 94  
Range % 87-97  
Recovery
The recovery of human CYP27B1 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type Average % Recovery Range  
Serum (n=5) 90 85-96  
EDTA plasma (n=4) 92 87-96  
             
             
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
pg/ml OD1 OD2 Average Corrected  
2000 2.472 2.372 2.422 2.248  
1000 1.715 1.615 1.665 1.491  
500 1.088 1.048 1.068 0.894  
250 0.711 0.701 0.706 0.532  
125 0.404 0.394 0.399 0.225  
62.5 0.310 0.300 0.305 0.131  
31.25 0.249 0.239 0.244 0.070  
0 0.175 0.173 0.174    
Storage Store at 2-8°C. Please refer to protocol.