Product Details

Target Name arginase, type II
Alternative Names ARG2 ELISA Kit; ARGI2_HUMAN ELISA Kit; Arginase II mitochondrial ELISA Kit; Arginase type II ELISA Kit; Arginase-2 ELISA Kit; arginase-2, mitochondrial ELISA Kit; Kidney arginase ELISA Kit; Kidney type arginase ELISA Kit; Kidney-type arginase ELISA Kit; L arginine amidinohydrolase ELISA Kit; L arginine ureahydrolase ELISA Kit; mitochondrial ELISA Kit; Non hepatic arginase ELISA Kit; Non-hepatic arginase ELISA Kit; Nonhepatic arginase ELISA Kit; Type II arginase ELISA Kit
Abbreviation ARG2
Species Homo sapiens (Human)
Sample Types serum, plasma, cell culture supernates, urine, cerebrospinal fluid (CSF)
Detection Range 0.47 mU/mL-30 mU/mL
Sensitivity 0.12 mU/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Metabolism
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%   
Three samples of known concentration were tested twenty times on one plate to assess.  
Inter-assay Precision (Precision between assays): CV%   
Three samples of known concentration were tested in twenty assays to assess.    
             
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human ARG2 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
  Sample Serum(n=4)  
1:1 Average % 87  
Range % 83-91  
1:2 Average % 104  
Range % 101-107  
1:4 Average % 96  
Range % 92-99  
1:8 Average % 85  
Range % 82-89  
Recovery
The recovery of human ARG2 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type Average % Recovery Range  
Serum (n=5) 94 91-98  
EDTA plasma (n=4) 90 86-97  
             
             
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
mU/ml OD1 OD2 Average Corrected  
30 2.327 2.298 2.313 2.146  
15 1.820 1.801 1.811 1.644  
7.5 1.393 1.378 1.386 1.219  
3.75 0.979 0.998 0.989 0.822  
1.87 0.564 0.610 0.587 0.420  
0.94 0.396 0.376 0.386 0.219  
0.47 0.273 0.295 0.284 0.117  
0 0.166 0.168 0.167    
Storage Store at 2-8°C. Please refer to protocol.
Lead Time 3-5 working days