18 years of successful growth

Human Collagen alpha-1(X) chain (COL10A1) ELISA kit

Product Details

Target Name collagen, type X, alpha 1
Alternative Names COAA1_HUMAN ELISA Kit; Col10a 1 ELISA Kit; COL10A1 ELISA Kit; Collagen alpha 1(X) chain ELISA Kit; Collagen alpha-1(X) chain ELISA Kit; Collagen type X alpha 1 (Schmid metaphyseal chondrodysplasia) ELISA Kit; Collagen type X alpha 1 ELISA Kit; Collagen X alpha 1 polypeptide ELISA Kit; CollagenX ELISA Kit; fa66d11 ELISA Kit; fb10c08 ELISA Kit; OTTHUMP00000040411 ELISA Kit; Procollagen type X alpha 1 ELISA Kit; Schmid metaphyseal chondrodysplasia ELISA Kit; wu:fa66d11 ELISA Kit; wu:fb10c08 ELISA Kit
Abbreviation COL10A1
Species Homo sapiens (Human)
Sample Types serum, plasma, tissue homogenates, cell lysates
Detection Range 0.312 ng/mL-20 ng/mL
Sensitivity 0.078 ng/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Signal Transduction
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%   
Three samples of known concentration were tested twenty times on one plate to assess.  
Inter-assay Precision (Precision between assays): CV%   
Three samples of known concentration were tested in twenty assays to assess.    
             
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human COL10A1 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
  Sample Serum(n=4)  
1:1 Average % 95  
Range % 90-106  
1:2 Average % 92  
Range % 88-95  
1:4 Average % 95  
Range % 90-101  
1:8 Average % 96  
Range % 91-103  
Recovery
The recovery of human COL10A1 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type Average % Recovery Range  
Serum (n=5) 96 92-101  
EDTA plasma (n=4) 92 87-99  
             
             
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
ng/ml OD1 OD2 Average Corrected  
20 3.039 3.052 3.046 2.909  
10 2.647 2.658 2.653 2.516  
5 1.898 2.040 1.969 1.832  
2.5 1.211 1.262 1.237 1.100  
1.25 0.708 0.663 0.686 0.549  
0.625 0.425 0.406 0.416 0.279  
0.312 0.242 0.257 0.250 0.113  
0 0.138 0.136 0.137    
Storage Store at 2-8°C. Please refer to protocol.
Lead Time 7-14 working days