Human Eotaxin 2/CCL24 ELISA Kit

Product Details

Target Name chemokine (C-C motif) ligand 24
Alternative Names C C motif chemokine 24 ELISA Kit; C-C motif chemokine 24 ELISA Kit; CCL24 ELISA Kit; CCL24_HUMAN ELISA Kit; Chemokine CC Motif Ligand 24 ELISA Kit; CK beta 6 ELISA Kit; CK-beta-6 ELISA Kit; Ckb6 ELISA Kit; Eosinophil chemotactic protein 2 ELISA Kit; Eotaxin-2 ELISA Kit; MPIF 2 ELISA Kit; MPIF-2 ELISA Kit; MPIF2 ELISA Kit; Myeloid progenitor inhibitory factor 2 ELISA Kit; SCYA24 ELISA Kit; Small inducible cytokine A24 ELISA Kit; Small inducible cytokine subfamily A (Cys-Cys) member 24 ELISA Kit; Small-inducible cytokine A24 ELISA Kit
Abbreviation CCL24
Species Homo sapiens (Human)
Sample Types serum, plasma, tissue homogenates
Detection Range 78.125 pg/mL-5000 pg/mL
Sensitivity 67.660 pg/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Immunology
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%        
Three samples of known concentration were tested twenty times on one plate to assess.    
Inter-assay Precision (Precision between assays): CV%        
Three samples of known concentration were tested in twenty assays to assess.      
               
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human eotaxin 2/CCL24 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.  
  Sample Serum(n=4)    
1:1 Average % 95    
Range % 90-100    
1:2 Average % 89    
Range % 85-94    
1:4 Average % 96    
Range % 91-101    
1:8 Average % 104    
Range % 100-108    
Recovery
The recovery of human eotaxin 2/CCL24 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.  
 
Sample Type Average % Recovery Range    
Serum (n=5) 99 95-103    
EDTA plasma (n=4) 86 83-89    
               
               
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.  
 
pg/ml OD1 OD2 Average Corrected    
5000 2.152 2.204 2.178 2.065    
2500 1.796 1.735 1.766 1.653    
1250 1.182 1.205 1.194 1.081    
625 0.765 0.798 0.782 0.669    
312.5 0.453 0.468 0.461 0.348    
156.25 0.328 0.331 0.330 0.217    
78.125 0.251 0.240 0.246 0.133    
0 0.115 0.110 0.113      
Storage Store at 2-8°C. Please refer to protocol.
Lead Time 3-5 working days