18 years of successful growth

Human Ficolin-2(FCN2) ELISA kit

Product Details

Target Name ficolin (collagen/fibrinogen domain containing lectin) 2 (hucolin)
Alternative Names 37 kDa elastin-binding protein ELISA Kit; Collagen/fibrinogen domain containing protein 2 ELISA Kit; Collagen/fibrinogen domain-containing protein 2 ELISA Kit; EBP 37 ELISA Kit; EBP-37 ELISA Kit; EBP37 ELISA Kit; FCN 2 ELISA Kit; Fcn2 ELISA Kit; FCN2_HUMAN ELISA Kit; FCNL ELISA Kit; Ficolin (collagen/fibrinogen domain containing lectin) 2 (hucolin) ELISA Kit; Ficolin B ELISA Kit; Ficolin beta ELISA Kit; Ficolin-2 ELISA Kit; Ficolin-B ELISA Kit; Ficolin-beta ELISA Kit; Ficolin2 ELISA Kit; Hucolin ELISA Kit; L ficolin ELISA Kit; L-ficolin ELISA Kit; OTTHUMP00000022518 ELISA Kit; P35 ELISA Kit; RP11 263F14.2 ELISA Kit; Serum lectin p35 ELISA Kit
Abbreviation FCN2
Species Homo sapiens (Human)
Sample Types serum, plasma, tissue homogenates
Detection Range 0.25 ng/mL-16 ng/mL
Sensitivity 0.06 ng/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Immunology
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%   
Three samples of known concentration were tested twenty times on one plate to assess.  
Inter-assay Precision (Precision between assays): CV%   
Three samples of known concentration were tested in twenty assays to assess.    
             
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human FCN2 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
  Sample Serum(n=4)  
1:1000 Average % 86  
Range % 80-94  
1:2000 Average % 95  
Range % 91-99  
1:4000 Average % 108  
Range % 98-115  
1:8000 Average % 95  
Range % 89-98  
Recovery
The recovery of human FCN2 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type Average % Recovery Range  
Serum (n=5) 95 89-98  
EDTA plasma (n=4) 101 98-106  
             
             
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
ng/ml OD1 OD2 Average Corrected  
16 2.007 1.900 1.954 1.844  
8 1.526 1.443 1.485 1.375  
4 1.097 1.009 1.053 0.943  
2 0.645 0.621 0.633 0.523  
1 0.429 0.420 0.425 0.315  
0.5 0.264 0.254 0.259 0.149  
0.25 0.190 0.186 0.188 0.078  
0 0.110 0.109 0.110    
Storage Store at 2-8°C. Please refer to protocol.
Lead Time 5-7 working days
Download X-Ray Brochures