18 years of successful growth

Human glutamic acid decarboxylase,GAD ELISA Kit

Product Details

Target Name glutamate decarboxylase 1 (brain, 67kDa)
Alternative Names 67 kDa glutamic acid decarboxylase ELISA Kit; CPSQ1 ELISA Kit; DCE1 ELISA Kit; DCE1_HUMAN ELISA Kit; EC 4.1.1.15 ELISA Kit; FLJ45882 ELISA Kit; GAD 1 ELISA Kit; GAD 67 ELISA Kit; GAD ELISA Kit; GAD-67 ELISA Kit; GAD1 ELISA Kit; Glutamate decarboxylase 1 (brain; 67kDa) ELISA Kit; Glutamate decarboxylase 1 ELISA Kit; Glutamate decarboxylase 1 brain 67kD ELISA Kit; Glutamate decarboxylase 1 brain 67kDa ELISA Kit; Glutamate decarboxylase 67 kDa isoform ELISA Kit; Glutamate decarboxylase; brain; 67-KD ELISA Kit; OTTHUMP00000041054 ELISA Kit; OTTHUMP00000041055 ELISA Kit; OTTHUMP00000228572 ELISA Kit; SCP ELISA Kit
Abbreviation GAD1
Species Homo sapiens (Human)
Sample Types serum, plasma, tissue homogenates
Detection Range 0.312 ng/mL-20 ng/mL
Sensitivity 0.078 ng/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Neuroscience
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%   
Three samples of known concentration were tested twenty times on one plate to assess.  
Inter-assay Precision (Precision between assays): CV%   
Three samples of known concentration were tested in twenty assays to assess.    
             
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human GAD in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
  Sample Serum(n=4)  
1:1 Average % 93  
Range % 90-98  
1:2 Average % 86  
Range % 82-93  
1:4 Average % 95  
Range % 90-100  
1:8 Average % 92  
Range % 87-96  
Recovery
The recovery of human GAD spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type Average % Recovery Range  
Serum (n=5) 90 85-95  
EDTA plasma (n=4) 99 95-102  
             
             
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
ng/ml OD1 OD2 Average Corrected  
20 2.512 2.458 2.485 2.357  
10 1.678 1.788 1.733 1.605  
5 1.105 1.190 1.148 1.020  
2.5 0.685 0.710 0.698 0.570  
1.25 0.474 0.465 0.470 0.342  
0.625 0.340 0.321 0.331 0.203  
0.312 0.240 0.221 0.231 0.103  
0 0.125 0.131 0.128    
Storage Store at 2-8°C. Please refer to protocol.
Lead Time 3-5 working days