Human low-density lipoprotein-receptor-related protein,LRP-6 ELISA Kit

Product Details

Target Name low density lipoprotein receptor-related protein 6
Alternative Names ADCAD2 ELISA Kit; C030016K15Rik ELISA Kit; Cd ELISA Kit; FLJ90062 ELISA Kit; FLJ90421 ELISA Kit; LDL receptor related protein 6 ELISA Kit; Low density lipoprotein receptor related protein 6 ELISA Kit; Low-density lipoprotein receptor-related protein 6 ELISA Kit; LRP-6 ELISA Kit; LRP6 ELISA Kit; LRP6_HUMAN ELISA Kit; OTTHUMP00000238979 ELISA Kit; OTTHUMP00000238980 ELISA Kit; OTTHUMP00000238982 ELISA Kit; STHAG7 ELISA Kit
Abbreviation LRP6
Species Homo sapiens (Human)
Sample Types serum, tissue homogenates, cell lysates
Detection Range 0.312 ng/mL-20 ng/mL
Sensitivity 0.078 ng/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Signal Transduction
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%   
Three samples of known concentration were tested twenty times on one plate to assess.  
Inter-assay Precision (Precision between assays): CV%   
Three samples of known concentration were tested in twenty assays to assess.    
             
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of Human LRP-6 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
  Sample Serum(n=4)  
1:1 Average % 88  
Range % 83-91  
1:2 Average % 92  
Range % 86-97  
1:4 Average % 94  
Range % 88-98  
1:8 Average % 102  
Range % 98-104  
Recovery
The recovery of Human LRP-6 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type Average % Recovery Range  
Serum (n=5) 94 88-98  
EDTA plasma (n=4) 86 82-92  
             
             
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
ng/ml OD1 OD2 Average Corrected  
20 2.625 2.729 2.677 2.494  
10 2.146 2.047 2.097 1.914  
5 1.444 1.463 1.454 1.271  
2.5 0.912 0.928 0.920 0.737  
1.25 0.597 0.586 0.592 0.409  
0.625 0.401 0.416 0.409 0.226  
0.312 0.287 0.275 0.281 0.098  
0 0.182 0.183 0.183    
Storage Store at 2-8°C. Please refer to protocol.