Human Matrix Metalloproteinase 12(MMP-12) ELISA Kit

Product Details

Target Name matrix metallopeptidase 12 (macrophage elastase)
Alternative Names EC 3.4.24.65 ELISA Kit; HME ELISA Kit; Macrophage elastase ELISA Kit; Macrophage metalloelastase ELISA Kit; Macrophage metaloelastase ELISA Kit; Matrix metallopeptidase 12 (macrophage elastase) ELISA Kit; Matrix metalloprotease 12 ELISA Kit; Matrix metalloproteinase-12 ELISA Kit; ME ELISA Kit; MGC138506 ELISA Kit; MME ELISA Kit; MMP 12 ELISA Kit; MMP-12 ELISA Kit; Mmp12 ELISA Kit; MMP12_HUMAN ELISA Kit
Abbreviation MMP12
Species Homo sapiens (Human)
Sample Types serum, plasma, tissue homogenates
Detection Range 0.625 ng/mL-40 ng/mL
Sensitivity 0.156 ng/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Cancer
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%   
Three samples of known concentration were tested twenty times on one plate to assess.  
Inter-assay Precision (Precision between assays): CV%   
Three samples of known concentration were tested in twenty assays to assess.    
             
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human MMP-12 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
  Sample Serum(n=4)  
1:1 Average % 97  
Range % 92-104  
1:2 Average % 89  
Range % 81-96  
1:4 Average % 98  
Range % 94-106  
1:8 Average % 95  
Range % 89-99  
Recovery
The recovery of human MMP-12 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type Average % Recovery Range  
Serum (n=5) 93 88-98  
EDTA plasma (n=4) 97 92-107  
             
             
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
ng/ml OD1 OD2 Average Corrected  
40 2.695 2.594 2.645 2.514  
20 2.132 2.031 2.082 1.951  
10 1.423 1.444 1.434 1.303  
5 0.813 0.802 0.808 0.677  
2.5 0.455 0.445 0.450 0.319  
1.25 0.290 0.301 0.296 0.165  
0.625 0.220 0.214 0.217 0.086  
0 0.131 0.130 0.131    
Storage Store at 2-8°C. Please refer to protocol.