18 years of successful growth

Human omentin ELISA Kit

Product Details

Target Name intelectin 1 (galactofuranose binding)
Alternative Names Endothelial lectin HL 1 ELISA Kit; Endothelial lectin HL-1 ELISA Kit; FLJ20022 ELISA Kit; Galactofuranose binding lectin ELISA Kit; Galactofuranose-binding lectin ELISA Kit; hIntL ELISA Kit; HL 1 ELISA Kit; HL-1 ELISA Kit; HL1 ELISA Kit; Intelectin ELISA Kit; Intelectin 1 ELISA Kit; Intelectin-1 ELISA Kit; Intestinal lactoferrin receptor ELISA Kit; INTL ELISA Kit; ITLN ELISA Kit; ITLN-1 ELISA Kit; ITLN1 ELISA Kit; ITLN1_HUMAN ELISA Kit; LFR ELISA Kit; Omentin ELISA Kit; OTTHUMP00000027882 ELISA Kit
Abbreviation ITLN1
Species Homo sapiens (Human)
Sample Types serum, plasma, cell culture supernates, tissue homogenates
Detection Range 1.56 pg/mL-100 pg/mL
Sensitivity 0.39 pg/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Cardiovascular
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%   
Three samples of known concentration were tested twenty times on one plate to assess.  
Inter-assay Precision (Precision between assays): CV%   
Three samples of known concentration were tested in twenty assays to assess.    
             
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human omentin in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
  Sample Serum(n=4)  
1:200 Average % 85  
Range % 80-92  
1:400 Average % 96  
Range % 92-100  
1:800 Average % 102  
Range % 93-110  
1:1600 Average % 95  
Range % 82-99  
Recovery
The recovery of human omentin spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type Average % Recovery Range  
Serum (n=5) 91 80-100  
EDTA plasma (n=4) 96 90-105  
             
             
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
pg/ml OD1 OD2 Average Corrected  
100 2.349 2.192 2.271 2.166  
50 1.869 1.806 1.838 1.733  
25 1.354 1.294 1.324 1.219  
12.5 0.923 0.914 0.919 0.814  
6.25 0.592 0.559 0.576 0.471  
3.12 0.322 0.299 0.311 0.206  
1.56 0.213 0.209 0.211 0.106  
0 0.109 0.101 0.105    
Storage Store at 2-8°C. Please refer to protocol.
Lead Time 3-5 working days
Download X-Ray Brochures