Human Phospholipidase A2,PLA2 ELISA Kit

Product Details

Target Name phospholipase A2, group IB (pancreas)
Alternative Names Group IB phospholipase A2 ELISA Kit; Group IB secretory phospholipase A2 ELISA Kit; MGC119834 ELISA Kit; MGC119835 ELISA Kit; PA21B_HUMAN ELISA Kit; Phosphatidylcholine 2 acylhydrolase ELISA Kit; Phosphatidylcholine 2-acylhydrolase 1B ELISA Kit; Phospholipase A2 ELISA Kit; Phospholipase A2 group IB (pancreas) ELISA Kit; Phospholipase A2 group IB ELISA Kit; PLA 2 ELISA Kit; PLA2 ELISA Kit; PLA2A ELISA Kit; PLA2G1B ELISA Kit; PPLA 2 ELISA Kit; PPLA2 ELISA Kit; sPLA2IB ELISA Kit
Abbreviation PLA2G1B
Species Homo sapiens (Human)
Sample Types serum, plasma, tissue homogenates
Detection Range 0.45 ng/mL-30 ng/mL
Sensitivity 0.11 ng/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Metabolism
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%   
Three samples of known concentration were tested twenty times on one plate to assess.  
Inter-assay Precision (Precision between assays): CV%   
Three samples of known concentration were tested in twenty assays to assess.    
             
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human PLA2 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
  Sample Serum(n=4)  
1:5 Average % 93  
Range % 88-97  
1:10 Average % 104  
Range % 100-109  
1:20 Average % 95  
Range % 91-99  
1:40 Average % 85  
Range % 81-92  
Recovery
The recovery of human PLA2 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type Average % Recovery Range  
Serum (n=5) 92 88-96  
EDTA plasma (n=4) 97 93-101  
             
             
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
ng/ml OD1 OD2 Average Corrected  
30 2.534 2.452 2.493 2.384  
15 1.652 1.576 1.614 1.505  
7.5 1.066 1.004 1.035 0.926  
3.75 0.578 0.561 0.570 0.461  
1.8 0.387 0.401 0.394 0.285  
0.9 0.238 0.245 0.242 0.133  
0.45 0.179 0.187 0.183 0.074  
0 0.105 0.112 0.109    