Human secretory leukocyte protease inhibitor,SLPI ELISA Kit

Product Details

Target Name secretory leukocyte peptidase inhibitor
Alternative Names ALK 1 ELISA Kit; ALK1 ELISA Kit; ALP ELISA Kit; Antileukoprotease ELISA Kit; Antileukoproteinase 1 ELISA Kit; Antileukoproteinase ELISA Kit; Antileukoproteinase1 ELISA Kit; BLPI ELISA Kit; Human seminal protease inhibitor ELISA Kit; HUSI 1 ELISA Kit; HUSI ELISA Kit; HUSI I ELISA Kit; HUSI-1 ELISA Kit; MPI ELISA Kit; Mucus proteinase inhibitor ELISA Kit; Protease inhibitor WAP4 ELISA Kit; Secretory leukocyte peptidase inhibitor ELISA Kit; Secretory leukocyte protease inhibitor ELISA Kit; Seminal proteinase inhibitor ELISA Kit; SLPI ELISA Kit; SLPI_HUMAN ELISA Kit; WAP four disulfide core domain protein 4 ELISA Kit; WAP four-disulfide core domain protein 4 ELISA Kit; WAP4 ELISA Kit; WFDC4 ELISA Kit
Abbreviation SLPI
Species Homo sapiens (Human)
Sample Types serum, plasma, tissue homogenates
Detection Range 31.25 pg/mL-2000 pg/mL
Sensitivity 7.81 pg/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Cardiovascular
Precision
Intra-assay Precision (Precision within an assay): CV%        
Three samples of known concentration were tested twenty times on one plate to assess.    
Inter-assay Precision (Precision between assays): CV%        
Three samples of known concentration were tested in twenty assays to assess.      
               
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human SLPI in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.  
  Sample Serum(n=4)    
1:100 Average % 96    
Range % 91-102    
1:200 Average % 92    
Range % 85-98    
1:400 Average % 89    
Range % 83-94    
1:800 Average % 91    
Range % 87-96    
Recovery
The recovery of human SLPI spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.  
 
Sample Type Average % Recovery Range    
Serum (n=5) 97 93-102    
EDTA plasma (n=4) 86 81-92    
               
               
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.  
 
pg/ml OD1 OD2 Average Corrected    
2000 2.637 2.531 2.584 2.490    
1000 2.124 2.205 2.165 2.071    
500 1.454 1.557 1.506 1.412    
250 0.936 0.923 0.930 0.836    
125 0.448 0.478 0.463 0.369    
62.5 0.283 0.274 0.279 0.185    
31.25 0.156 0.163 0.160 0.066    
Storage Store at 2-8°C. Please refer to protocol.
Lead Time 3-5 working days