18 years of successful growth

Human tumor necrosis factor receptor-associated factor 6,TRAF6 ELISA kit

Product Details

Target Name TNF receptor-associated factor 6
Alternative Names E3 ubiquitin-protein ligase TRAF6 ELISA Kit; Interleukin 1 signal transducer ELISA Kit; Interleukin-1 signal transducer ELISA Kit; MGC 3310 ELISA Kit; MGC:3310 ELISA Kit; MGC3310 ELISA Kit; OTTHUMP00000232772 ELISA Kit; OTTHUMP00000232773 ELISA Kit; RING finger protein 85 ELISA Kit; RNF 85 ELISA Kit; RNF85 ELISA Kit; TNF receptor associated factor 6 ELISA Kit; TNF receptor-associated factor 6 ELISA Kit; TNF receptor-associated factor 6; E3 ubiquitin protein ligase ELISA Kit; TRAF 6 ELISA Kit; Traf6 ELISA Kit; TRAF6_HUMAN ELISA Kit
Abbreviation TRAF6
Species Homo sapiens (Human)
Sample Types serum, plasma, tissue homogenates, cell lysates
Detection Range 39 pg/mL-2500 pg/mL
Sensitivity 9.8 pg/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Signal Transduction
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%   
Three samples of known concentration were tested twenty times on one plate to assess.  
Inter-assay Precision (Precision between assays): CV%   
Three samples of known concentration were tested in twenty assays to assess.    
             
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human TRAF6 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
  Sample Serum(n=4)  
1:1 Average % 92  
Range % 87-99  
1:2 Average % 103  
Range % 99-104  
1:4 Average % 93  
Range % 87-98  
1:8 Average % 91  
Range % 89-97  
Recovery
The recovery of human TRAF6 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type Average % Recovery Range  
Serum (n=5) 95 92-99  
EDTA plasma (n=4) 90 86-93  
             
             
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
pg/ml OD1 OD2 Average Corrected  
2500 2.444 2.246 2.345 2.239  
1250 1.937 1.780 1.859 1.753  
625 1.432 1.544 1.488 1.382  
312.5 1.078 1.159 1.119 1.013  
156 0.726 0.760 0.743 0.637  
78 0.376 0.358 0.367 0.261  
39 0.261 0.244 0.253 0.147  
0 0.103 0.109 0.106    
Storage Store at 2-8°C. Please refer to protocol.
Lead Time 3-5 working days
Download X-Ray Brochures