18 years of successful growth

Mouse adrenocorticotropic hormone (ACTH) ELISA kit

Product Details

Target Name adrenocorticotropic hormone (ACTH)
Abbreviation ACTH
Species Mus musculus (Mouse)
Sample Types serum, plasma
Detection Range 18.75 pg/mL-1200 pg/mL
Sensitivity 4.68 pg/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Metabolism
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%   
Three samples of known concentration were tested twenty times on one plate to assess.  
Inter-assay Precision (Precision between assays): CV%   
Three samples of known concentration were tested in twenty assays to assess.    
             
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of mouse ACTH in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
  Sample Serum(n=4)  
1:1 Average % 92  
Range % 80-105  
1:2 Average % 96  
Range % 85-104  
1:4 Average % 98  
Range % 92-104  
1:8 Average % 96  
Range % 82-110  
Recovery
The recovery of mouse ACTH spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type Average % Recovery Range  
Serum (n=5) 95 84-102  
EDTA plasma (n=4) 98 90-105  
             
             
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
pg/ml OD1 OD2 Average Corrected  
1200 3.134 3.126 3.130 2.988  
600 2.537 2.532 2.535 2.393  
300 1.754 1.765 1.760 1.618  
150 1.053 1.050 1.052 0.910  
75 0.651 0.647 0.649 0.507  
37.5 0.443 0.435 0.439 0.297  
18.75 0.273 0.265 0.269 0.127  
0 0.140 0.144 0.142    
Storage Store at 2-8°C. Please refer to protocol.
Lead Time 3-5 working days