Product Details

Target Name angiotensinogen (serpin peptidase inhibitor, clade A, member 8)
Alternative Names Agt ELISA kit; Serpina8Angiotensinogen ELISA kit; Serpin A8) [Cleaved into: Angiotensin-1 ELISA kit; Angiotensin 1-10 ELISA kit; Angiotensin I ELISA kit; Ang I); Angiotensin-2 ELISA kit; Angiotensin 1-8 ELISA kit; Angiotensin II ELISA kit; Ang II); Angiotensin-3 ELISA kit; Angiotensin 2-8 ELISA kit; Angiotensin III ELISA kit; Ang III ELISA kit; Des-Asp[1]-angiotensin II); Angiotensin-4 ELISA kit; Angiotensin 3-8 ELISA kit; Angiotensin IV ELISA kit; Ang IV); Angiotensin 1-9; Angiotensin 1-7; Angiotensin 1-5; Angiotensin 1-4] ELISA kit
Abbreviation AGT
Species Mus musculus (Mouse)
Sample Types serum, plasma, cell culture supernates, tissue homogenates
Detection Range 3.12 ng/mL-200 ng/mL
Sensitivity 0.78 ng/mL
Assay Time 1-5h
Sample Volume 50-100ul
Detection Wavelength 450 nm
Research Area Cardiovascular
Assay Principle quantitative
Measurement Sandwich
Precision
Intra-assay Precision (Precision within an assay): CV%        
Three samples of known concentration were tested twenty times on one plate to assess.    
Inter-assay Precision (Precision between assays): CV%        
Three samples of known concentration were tested in twenty assays to assess.      
               
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of mouse aGT in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.  
  Sample Serum(n=4)    
1:100 Average % 97    
Range % 92-101    
1:200 Average % 102    
Range % 98-110    
1:400 Average % 91    
Range % 86-95    
1:800 Average % 93    
Range % 89-97    
Recovery
The recovery of mouse aGT spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.  
 
Sample Type Average % Recovery Range    
Serum (n=5) 89 85-93    
EDTA plasma (n=4) 84 80-88    
               
               
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.  
 
ng/ml OD1 OD2 Average Corrected    
200 2.584 2.445 2.515 2.410    
100 1.692 1.573 1.633 1.528    
50 0.951 1.016 0.984 0.879    
25 0.647 0.662 0.655 0.550    
12.5 0.359 0.339 0.349 0.244    
6.25 0.236 0.221 0.229 0.124    
3.12 0.183 0.187 0.185 0.080    
0 0.102 0.108 0.105      
Storage Store at 2-8°C. Please refer to protocol.
Lead Time 3-5 working days